First published: Tue Mar 26 2019(Updated: )
An unauthenticated remote attacker may be able to execute commands to view wireless account credentials that are stored in cleartext on Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16, which may allow an attacker to gain access the host network. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.
Credit: ics-cert@hq.dhs.gov
Affected Software | Affected Version | How to fix |
---|---|---|
Baxter Sigma Spectrum Infusion System | =6.05 | |
Baxter Sigma Spectrum Infusion System | ||
Baxter Wireless Battery Module | =16 |
Sign up to SecAlerts for real-time vulnerability data matched to your software, aggregated from hundreds of sources.
The severity of CVE-2014-5433 is classified as high due to the potential for unauthenticated remote command execution.
To fix CVE-2014-5433, upgrade the Baxter SIGMA Spectrum Infusion System to a version that is not vulnerable, beyond version 6.05.
CVE-2014-5433 affects Baxter SIGMA Spectrum Infusion System version 6.05 and Wireless Battery Module version 16.
Exploitation of CVE-2014-5433 could allow an attacker to view sensitive wireless account credentials stored in cleartext.
Currently, the recommended solution for CVE-2014-5433 is to upgrade to a secure version, as no specific workarounds have been published.