First published: Mon Jun 29 2020(Updated: )
Baxter ExactaMix EM 2400 Versions 1.10, 1.11, and 1.13 and ExactaMix EM1200 Versions 1.1, 1.2, and 1.4 does not restrict non administrative users from gaining access to the operating system and editing the application startup script. Successful exploitation of this vulnerability may allow an attacker to alter the startup script as the limited-access user.
Credit: ics-cert@hq.dhs.gov
Affected Software | Affected Version | How to fix |
---|---|---|
Baxter EM2400 | =1.10 | |
Baxter EM2400 | =1.11 | |
Baxter EM2400 | =1.13 | |
Baxter | ||
Baxter Em1200 | =1.1 | |
Baxter Em1200 | =1.2 | |
Baxter Em1200 | =1.4 | |
Baxter Em1200 Firmware |
Sign up to SecAlerts for real-time vulnerability data matched to your software, aggregated from hundreds of sources.
CVE-2020-12020 is classified with a high severity due to potential unauthorized access to the operating system.
To mitigate CVE-2020-12020, upgrade to a patched version of the firmware provided by Baxter.
CVE-2020-12020 affects Baxter ExactaMix EM 2400 firmware versions 1.10, 1.11, and 1.13, and ExactaMix EM1200 firmware versions 1.1, 1.2, and 1.4.
Exploitation of CVE-2020-12020 may allow non-administrative users to access and modify the application startup script.
The responsibility for addressing CVE-2020-12020 lies with the device owners, who should implement firmware updates issued by Baxter.