CVE-2024-1275: Vulnerability in Baxter Welch Allyn Connex Spot Monitor

Published May 31, 2024
·
Updated

Use of Default Cryptographic Key vulnerability in Baxter Welch Allyn Connex Spot Monitor may allow Configuration/Environment Manipulation.This issue affects Welch Allyn Connex Spot Monitor in all versions prior to 1.52.

Affected Software

1 affected component
Baxter Welch Allyn Connex Spot Monitor<1.52

Remediation

Information

Baxter has released a software update for all impacted devices and software to address this vulnerability. A new version of the product that mitigates the vulnerability is available as follows: * Welch Allyn Connex Spot Monitor: Version 1.52.01 (available October 16, 2023) Baxter recommends users upgrade to the latest versions of their products. Information on how to update products to their new versions can be found on the Baxter disclosure page https://www.baxter.com/product-security  or the Hillrom disclosure page https://www.hillrom.com/en/responsible-disclosures/ . Baxter recommends the following workarounds to help reduce risk: * Apply proper network and physical security controls. * Ensure a unique encryption key is configured and applied to the product (as described in the Connex Spot Monitor Service Manual).

Event History

May 31, 2024
CVE Published
via MITRE·05:23 PM
Data Sourced
via MITRE·05:23 PM
RemedyDescriptionWeakness
Data Sourced
via NVD·06:15 PM
DescriptionSeverityWeakness
Free Weekly Intel

Don't miss critical vulnerabilities

Join thousands of security professionals who receive our weekly digest of trending CVEs, zero-days, and exploited vulnerabilities.

No spam. Unsubscribe anytime.

Frequently Asked Questions

1

What is the severity of CVE-2024-1275?

CVE-2024-1275 is classified with a severity level that indicates a significant risk due to the potential for configuration and environment manipulation.

2

How do I fix CVE-2024-1275?

To mitigate CVE-2024-1275, upgrade the Welch Allyn Connex Spot Monitor to version 1.52 or later.

3

Which products are affected by CVE-2024-1275?

CVE-2024-1275 affects all versions of the Welch Allyn Connex Spot Monitor prior to version 1.52.

4

What type of vulnerability is CVE-2024-1275?

CVE-2024-1275 is a use of default cryptographic key vulnerability.

5

Can CVE-2024-1275 lead to unauthorized access?

Yes, CVE-2024-1275 may allow attackers to manipulate configurations and environments, potentially leading to unauthorized access.

Contact

SecAlerts Pty Ltd.
132 Wickham Terrace
Fortitude Valley,
QLD 4006, Australia
info@secalerts.co
By using SecAlerts services, you agree to our services end-user license agreement. This website is safeguarded by reCAPTCHA and governed by the Google Privacy Policy and Terms of Service. All names, logos, and brands of products are owned by their respective owners, and any usage of these names, logos, and brands for identification purposes only does not imply endorsement. If you possess any content that requires removal, please get in touch with us.