First published: Fri May 17 2024(Updated: )
This vulnerability allows remote attackers to execute arbitrary code on affected installations of Sante PACS Server PG. Authentication is not required to exploit this vulnerability. The specific flaw exists within the implementation of the DICOM service, which listens on TCP port 11122 by default. When parsing the NAME element of the PATIENT record, the process does not properly validate a user-supplied string before using it to construct SQL queries. An attacker can leverage this vulnerability to execute code in the context of NETWORK SERVICE.
Affected Software | Affected Version | How to fix |
---|---|---|
Sante PACS Server PG |
Sign up to SecAlerts for real-time vulnerability data matched to your software, aggregated from hundreds of sources.
The severity of ZDI-24-468 is critical due to its ability to allow remote code execution without authentication.
To fix ZDI-24-468, users should apply the latest security patches provided by Sante for PACS Server PG.
Organizations using the Sante PACS Server PG are affected by ZDI-24-468.
Yes, ZDI-24-468 could lead to data breaches as it allows attackers to execute arbitrary code remotely.
The vulnerable component in ZDI-24-468 is the DICOM service implementation listening on TCP port 111.