An issue was discovered in Becton, Dickinson and Company (BD) Alaris 8015 Point of Care (PC) unit, Version 9.5 and prior versions, and Version 9.7, and 8000 PC unit. An unauthorized user with physical access to an affected Alaris PC unit may be able to obtain unencrypted wireless network authentication credentials and other sensitive technical data by disassembling the PC unit and accessing the device's flash memory. The Alaris 8015 PC unit, Version 9.7, and the 8000 PC unit store wireless network authentication credentials and other sensitive technical data on internal flash memory. Accessing the internal flash memory of the affected device would require special tools to extract data and carrying out this attack at a healthcare facility would increase the likelihood of detection.
An issue was discovered in Becton, Dickinson and Company (BD) Alaris 8015 Point of Care (PC) unit, Version 9.5 and prior versions, and Version 9.7. An unauthorized user with physical access to an Alaris 8015 PC unit may be able to obtain unencrypted wireless network authentication credentials and other sensitive technical data by disassembling an Alaris 8015 PC unit and accessing the device's flash memory. Older software versions of the Alaris 8015 PC unit, Version 9.5 and prior versions, store wireless network authentication credentials and other sensitive technical data on the affected device's removable flash memory. Being able to remove the flash memory from the affected device reduces the risk of detection, allowing an attacker to extract stored data at the attacker's convenience.
The configuration from the PCU can be modified without authentication using physical connection to the PCU.
The data flowing between the PCU and its modules is insecure. A threat actor with physical access could potentially read or modify data by attaching a specially crafted device while an infusion is running.
BD Alaris PC Unit, Model 8015, Versions 9.33.1 and earlier and BD Alaris Systems Manager, Versions 4.33 and earlier The affected products are vulnerable to a network session authentication vulnerability within the authentication process between specified versions of the BD Alaris PC Unit and the BD Alaris Systems Manager. If exploited, an attacker could perform a denial-of-service attack on the BD Alaris PC Unit by modifying the configuration headers of data in transit. A denial-of-service attack could lead to a drop in the wireless capability of the BD Alaris PC Unit, resulting in manual operation of the PC Unit.
The configuration from the PCU can be modified without authentication using physical connection to the PCU.