Where
-Infinity
0
Severity
6.4
CVSS:3.1/AV:A/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:L

In Patient Information Center iX (PICiX) Versions C.02 and C.03, PerformanceBridge Focal Point Version A.01, IntelliVue patient monitors MX100, MX400-MX550, MX750, MX850, and IntelliVue X3 Versions N and prior, the software does not check or incorrectly checks the revocation status of a certificate, which may cause it to use a compromised certificate.

1 / 2
Source: MITRE

Remedy

Philips released the following versions to remediate reported vulnerabilities: * Patient Information Center iX (PICiX) Version C.03 * PerformanceBridge Focal Point * IntelliVue Patient Monitors Versions N.00 and N.01 * IntelliVue Patient Monitors Version M.04: Contact a Philips service support team https://www.usa.philips.com/healthcare/solutions/customer-service-solutions for an upgrade path * Certificate revocation within the system was implemented for PIC iX and Performance Bridge FocalPoint in 2023. The implementation of the IntelliVue Patient Monitors will be completed in Q3 of 2024.
First published (updated )
Severity
6.5
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H

In Patient Information Center iX (PICiX) Versions C.02, C.03, the software parses a formatted message or structure but does not handle or incorrectly handles a length field that is inconsistent with the actual length of the associated data, causing the application on the surveillance station to restart.

1 / 2

Remedy

Philips released the following versions to remediate reported vulnerabilities: * Patient Information Center iX (PICiX) Version C.03 * Certificate revocation within the system was implemented for PIC iX.
First published (updated )
Severity
3.5
XSS
CVSS:3.1/AV:A/AC:L/PR:L/UI:N/S:U/C:L/I:N/A:N

In Patient Information Center iX (PICiX) Versions B.02, C.02, C.03, the software does not neutralize or incorrectly neutralizes user-controllable input before it is placed in output that is then used as a webpage and served to other users. Successful exploitation could lead to unauthorized access to patient data via a read-only web application.

1 / 2

Remedy

Philips released the following versions to remediate reported vulnerabilities: * Patient Information Center iX (PICiX) Version C.03 * Certificate revocation within the system was implemented for PIC iX.
First published (updated )
Severity
4.3
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:L

In Patient Information Center iX (PICiX) Versions C.02, C.03, PerformanceBridge Focal Point Version A.01, the product receives input that is expected to be well-formed (i.e., to comply with a certain syntax) but it does not validate or incorrectly validates that the input complies with the syntax, causing the certificate enrollment service to crash. It does not impact monitoring but prevents new devices from enrolling.

1 / 2

Remedy

Philips released the following versions to remediate reported vulnerabilities: * Patient Information Center iX (PICiX) Version C.03 * PerformanceBridge Focal Point * Certificate revocation within the system was implemented for PIC iX and Performance Bridge FocalPoint.
First published (updated )
Severity
6.5
Input Validation
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H

In IntelliVue patient monitors MX100, MX400-550, MX600, MX700, MX750, MX800, MX850, MP2-MP90, and IntelliVue X2 and X3 Versions N and prior, the product receives input or data but does not validate or incorrectly validates that the input has the properties required to process the data safely and correctly, which can induce a denial-of-service condition through a system restart.

1 / 2
Source: MITRE

Remedy

Philips released the following versions to remediate reported vulnerabilities: * IntelliVue Patient Monitors Versions N.00 and N.01 * IntelliVue Patient Monitors Version M.04: Contact a Philips service support team https://www.usa.philips.com/healthcare/solutions/customer-service-solutions for an upgrade path * Certificate revocation implementation of the IntelliVue Patient Monitors will be completed in Q3 of 2024.
First published (updated )
Severity
5.8
CVSS:3.1/AV:L/AC:L/PR:N/UI:R/S:C/C:L/I:L/A:N

In Patient Information Center iX (PICiX) Versions B.02, C.02, C.03, the software saves user-provided information into a comma-separated value (CSV) file, but it does not neutralize or incorrectly neutralizes special elements that could be interpreted as a command when the file is opened by spreadsheet software.

1 / 2

Remedy

Philips released the following versions to remediate reported vulnerabilities: * Patient Information Center iX (PICiX) Version C.03
First published (updated )
Severity
8.8
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H

In Patient Information Center iX (PICiX) Version B.02, C.02, C.03, and PerformanceBridge Focal Point Version A.01, when an actor claims to have a given identity, the software does not prove or insufficiently proves the claim is correct.

1 / 2

Remedy

Philips released the following versions to remediate reported vulnerabilities: * Patient Information Center iX (PICiX) Version C.03 * PerformanceBridge Focal Point * Certificate revocation within the system was implemented for PIC iX and Performance Bridge FocalPoint.
First published (updated )
Severity
6.1
CVSS:3.1/AV:L/AC:L/PR:L/UI:N/S:U/C:H/I:N/A:N

The use of a hard-coded cryptographic key significantly increases the possibility encrypted data may be recovered from the Patient Information Center iX (PIC iX) Versions B.02, C.02, and C.03.

First published (updated )
Severity
6.5
Input Validation
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:N

Patient Information Center iX (PIC iX) Versions C.02 and C.03 receives input or data, but does not validate or incorrectly validates that the input has the properties required to process the data safely and correctly.

First published (updated )
Severity
6.5
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N

The use of a broken or risky cryptographic algorithm is an unnecessary risk that may result in the exposure of sensitive information, which affects the communications between Patient Information Center iX (PIC iX) Versions C.02 and C.03 and Efficia CM Series Revisions A.01 to C.0x and 4.0.

First published (updated )
Severity
6.8
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H

In Patient Information Center iX (PICiX) Versions B.02, C.02, C.03, the product exposes a resource to the wrong control sphere, providing unintended actors with inappropriate access to the resource. The application on the surveillance station operates in kiosk mode, which is vulnerable to local breakouts that could allow an attacker with physical access to escape the restricted environment with limited privileges.

1 / 2

Remedy

Philips released the following versions to remediate reported vulnerabilities: * Patient Information Center iX (PICiX) Version C.03 * Certificate revocation within the system was implemented for PIC iX and Performance Bridge FocalPoint in 2023. The implementation of the IntelliVue Patient Monitors will be completed in Q3 of 2024.
First published (updated )

Contact

SecAlerts Pty Ltd.
132 Wickham Terrace
Fortitude Valley,
QLD 4006, Australia
info@secalerts.co
By using SecAlerts services, you agree to our services end-user license agreement. This website is safeguarded by reCAPTCHA and governed by the Google Privacy Policy and Terms of Service. All names, logos, and brands of products are owned by their respective owners, and any usage of these names, logos, and brands for identification purposes only does not imply endorsement. If you possess any content that requires removal, please get in touch with us.